PIXE characterization of tissues surrounding metallic prostheses coated with biological glasses
Creators
Description
Biological glasses can be used as coatings for metallic prostheses in order to prevent corrosion. According to their composition, these glasses have different properties. We studied, in vivo, two glasses referred to as BVA and BVH. They are used as coatings of Ti6Al4V metallic implant. BVA glass disappears after 3 months of implantation and is replaced by bone. Prostheses initially coated by this glass have a larger osseous contact perimeter compared to the uncoated prostheses. This ensures a better anchoring of the implant and limits the micro-motions which cause wear debris. BVH glass keeps a constant composition during implantation and it is used like a layer which isolates metal implant from biological environment. In order to characterize the bony environment surrounding implants, we have used PIXE and RBS methods. This paper shows results of the behavior of bony tissue under micro-beam, the quality tests of new bone which replaces the BVA glass coating and the evaluation of corrosion effects. Titanium release in bony tissues begins when the metal surface of the prosthesis is exposed to biological fluids. After a few months of implantation, the titanium contamination is stabilized and remains localized within the first tens of micrometers of surrounding bone
Additional details
Identifiers
- DOI
- 10.1016/S0168-583X(03)01785-3;
- arXiv
- arXiv:0808.1402v5;
- PII
- S0168583X03017853;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
- Journal Volume
- 215
- Journal Issue
- 1-2
- Journal Page Range
- p. 214-222
- ISSN
- 0168-583X
- CODEN
- NIMBEU
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 36012152
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S36: MATERIALS SCIENCE;
- Descriptors DEI
- ALUMINIUM COMPOUNDS; COATINGS; CORROSION; GLASS; IMPLANTS; MEASURING METHODS; PIXE ANALYSIS; PROSTHESES; RUTHERFORD BACKSCATTERING SPECTROSCOPY; TITANIUM; TITANIUM COMPOUNDS; VANADIUM COMPOUNDS
- Descriptors DEC
- CHEMICAL ANALYSIS; CHEMICAL REACTIONS; ELEMENTS; MEDICAL SUPPLIES; METALS; NONDESTRUCTIVE ANALYSIS; SPECTROSCOPY; TRANSITION ELEMENT COMPOUNDS; TRANSITION ELEMENTS; X-RAY EMISSION ANALYSIS
Optional Information
- Copyright
- Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.