Comparison of photon-counting to storage phosphor plate mammography using contrast-detail phantom analysis

  • 1. Physics Department, Royal Institute of Technology, AlbaNova University Center, SE-106 91 Stockholm (Sweden)
  • 2. Pruefstelle fuer Strahlenschutz, 47167 Duisburg, Nordrhein-Westfalen (Germany)

Description

Two digital mammography systems, one based on scanning photon counting silicon detectors and the other on storage phosphor plates (CR), have been compared in terms of image quality and dose. Sets of images of a contrast-detail phantom (CDMAM3.4) were acquired for each system and dose level. The images were evaluated in the disc diameter range 0.16-1 mm using a computer program (CDcom) and the results were fitted to a psychometric curve for each disc diameter. The contrast-detail curve was summarized into one single figure of merit, the image quality index, and the dose efficiency was calculated. The errors of the calculated parameters were assessed using statistical analysis. It was found that the scanning photon-counting system can achieve the same image quality as the storage phosphor plate (CR) system at 30-38% of the average glandular dose

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2007.06.115

Additional details

Identifiers

DOI
10.1016/j.nima.2007.06.115;
PII
S0168-9002(07)01332-0;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
580
Journal Issue
2
Journal Page Range
p. 1101-1104
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
3. international conference on imaging techniques in subatomic physics, astrophysics, medicine, biology and industry
Dates
27-30 Sep 2006
Place
Stockholm (Sweden)

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.