Assessment of Siemens plessey electronic personal dosimeter

  • 1. Ontario Hydro, Toronto, ON (Canada)
  • 2. Atomic Energy of Canada Ltd., Chalk River, ON (Canada). Chalk River Nuclear Labs.

Description

This report presents the results of a laboratory assessment of the performance of a new type of personal dosimeter. The Electronic Personal Dosimeter, or EPD, was developed jointly by the National Radiological Protection Board and Siemens Plessey Controls Limited, both of the United Kingdom. Twenty pre-production units of the EPD and a reader were purchased by Ontario Hydro for the assessment. The tests were conducted jointly by Ontario Hydro's Health and Safety Division and AECL Research's Chalk River Laboratories (CRL), with funding from the Candu Owner's Group. A total of 26 tests were conducted, divided between Ontario Hydro and AECL. The test results were compared with the relevant requirements of three standards. In general, the performance of the EPD was found to be quite acceptable. It met most of the relevant requirements of the three standards and most of the design specifications. However, the following deficiencies were found: slow response time; sensitivity to high-frequency EMF; poor resistance to dropping; and an alarm that is not loud enough. In addition, the response of the EPD to low-energy beta rays may be too low for some applications. There were serious problems with the reliability of operation of the pre production EPDs used in these tests. 9 refs., 34 tabs., 20 figs

Availability note (English)

MF available from INIS under the Report Number.

Files

27014011.pdf

Files (1.1 MB)

Name Size Download all
md5:79872e05a0257a019c1203cefafd2c80
1.1 MB Preview Download

Additional details

Publishing Information

Imprint Pagination
71 p.
Report number
AECL--10958

Optional Information

Secondary number(s)
COG--94-23.