Magnetic probe construction using thick-film technology
Creators
Description
Thick-film technology has been successfully adapted for the design and fabrication of magnetic probes of a new type suitable for use in the simultaneous ultrahigh vacuum and high temperature environment of a nuclear fusion device. The maximum usable temperature is expected to be around 900 C. This new probe has a specific sensitivity (coupling area per unit volume) an order of magnitude higher than a conventional coil. The new probe in one implementation is capable of simultaneously measuring magnetic field in three orthogonal directions about a single spatial point and in two frequency ranges. Low frequency coils have a measured coupling area of 296--323 cm2, and a frequency response of about 300 kHz. High frequency coils have a design coupling area of 12--15 cm2
Additional details
Identifiers
- DOI
- 10.1063/1.1384431;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 72
- Journal Issue
- 8
- Journal Page Range
- p. 3249-3259
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 32064691
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Descriptors DEI
- CONSTRUCTION; DESIGN; FABRICATION; FREQUENCY RANGE; IMPLEMENTATION; MAGNETIC FIELDS; MAGNETIC PROBES; PROBES; SENSITIVITY; ULTRAHIGH VACUUM
- Descriptors DEC
- PROBES
Optional Information
- Contract/Grant/Project number
- AC02-76-CH03073
- Notes
- Othernumber: RSINAK000072000008003249000001; 018108RSI
- Funding organization
- (US)