Estimating the error of CT-based measurements of aortic lumen volume used in endovascular planning
- 1. University of Salford, Salford (United Kingdom)
- 2. Liverpool Vascular & Endovascular Service, Liverpool (United Kingdom)
- 3. Royal Liverpool & Broadgreen University Hospital, Liverpool (United Kingdom)
Description
Introduction: Planning of endovascular sealing of abdominal aortic aneurysms requires measurement of the aortic lumen volume. The aim of this study was to investigate mathematically the effect of intra- and inter-observer variability error, as well as cardiac cycle-related variability, on these measurements. Methods: Mean (±2SD) intra- and inter-observer error in lumen measurements and mean (+2SD) cardiac cycle-related variability were obtained from published literature and added to the measurement of the flow lumen volume of a 57 mm abdominal aortic aneurysm to calculate average and extreme error possibilities. Results: The aneurysm volume was measured at 165 ml. The calculated possible mean measurement error due to cardiac cycle variation, intra- and inter-observer variability was +11.0%, resulting in a potential measurement of 183.1 ml. The calculated extreme errors were +24.3% (if 2SD of all errors were added to the mean) and +3.5% (if 2SD of all errors, except cardiac cycle, were subtracted from the mean), resulting in potential measurements of 170.8 ml and 205.1 ml, respectively. When considering the errors combined, the proportion of patients who may have volume measurement errors of up to ±2.5 ml, ±2.6 to ±5.0 ml and ±5.1 to ±7.5 ml were 18%, 17% and 15%, respectively. Conclusion: Measurement of CT-based aortic lumen volumes in abdominal aortic aneurysms is imprecise. This has practical implications for the planning and the performance of complex endovascular therapies. - Highlights: • CT based aortic volume measurements are prone to error. • Cardiac cycle variations are likely to be the greatest source of error. • Errors may have implications especially when planning treatment and being alerted to potential complications.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radi.2017.08.001Additional details
Identifiers
- DOI
- 10.1016/j.radi.2017.08.001;
- PII
- S1078-8174(17)30129-3;
Publishing Information
- Journal Title
- Radiography (London 1995)
- Journal Volume
- 23
- Journal Issue
- 4
- Journal Page Range
- p. 287-291
- ISSN
- 1078-8174
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49049801
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AORTA; COMPUTERIZED TOMOGRAPHY; ERRORS; PATIENTS; PERFORMANCE; THERAPY
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; MEDICINE; ORGANS; TOMOGRAPHY
Optional Information