Published 1982 | Version v1
Report Restricted

Pulse shape discrimination in a portable beta-gamma dose rate meter

Description

A portable beta-gamma dose rate instrument capable of tissue-equivalent dose rate measurements in mixed beta-gamma fields has been developed and tested. The detector is a 5 mg/cm2-thick plastic scintillator with a 4 mg/cm2-window and a 1 cm-thick lucite light pipe to achieve the backscatter geometry required for this application. Pulse shape discrimination techniques are used to minimize contributions from Cerenkov events within the lucite light pipe, photomultiplier tube glass, and direct interaction within the photomultiplier tube dynode structure. The response of the prototype instrument has been demonstrated to be independent of beta end-point energy to a precision of +- 5% from 150 keV to 2.9 MeV and 1 mRem/h to 200 Rem/h. Response to gamma radiation in the energy range >400 keV is also accurate, but low energy photon response (<100 keV) is considerably lower than expected

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02; 3 as DE83005745.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
EGG-M--25282

Conference

Title
IEEE nuclear science symposium.
Dates
20-22 Oct 1982.
Place
Washington, DC (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
15006956
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BETA DOSIMETRY; DOSE RATES; DOSE-RESPONSE RELATIONSHIPS; DOSEMETERS; GAMMA DOSIMETRY; LIGHT PIPES; PLASTIC SCINTILLATION DETECTOR; SPECIFICATIONS
Descriptors DEC
DOSIMETRY; MEASURING INSTRUMENTS; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS

Optional Information

Secondary number(s)
CONF-821011--42.