Detector ensemble with CsI(Tl) scintillator and PIN photodiode used in gamma radiation dosimetry

  • 1. Laboratory of Detectors and New Dosimetrical Methods, National Institute for Physics and Nuclear Engineering-Horia Hulubei, PO Box MG-6, RO - 76900 Bucharest - Magurele (Romania)

Description

In the paper the realization of a detector ensemble, the functional principle and performances are presented. The efficiency of detection obtained for gamma radiation with Eγ = 660 keV is ≥ 70 % for the crystal size of 25 mm in diameter and 35 mm in height. The energy resolution is higher than 6%. These values obtained with crystal coupled to PIN photodiode of type S, are equivalent with those obtained with the crystal coupled to the photomultiplier. The data were obtained using an acquisition program entitled 'Quantum MCA Leap' and an PCA-3-OXFORD multi-channel analyzer card mounted in a computer PC/AT-386. (authors)

Availability note (English)

Available from author(s) or from National Institute for Physics and Nuclear Engineering-Horia Hulubei, Str. Atomistilor No. 109, PO Box MG-6, RO-76900, Bucharest-Magurele (RO)
Part of:
NIPNE - Scientific Report 1996

Additional details

Publishing Information

Imprint Title
NIPNE-Scientific Report 1996
Imprint Pagination
286 p.
Journal Page Range
p. 222
Report number
NIPNE-AR--1997

Optional Information

Notes
Short communication