Aim: To evaluate the use of a phantom system to help teach the basic techniques of accurate CT-guided needle placement, thereby avoiding the risks associated with teaching on patients. Materials and Methods: Gelatine phantoms with five, 1.9 cm embedded spherical wooden targets were constructed. Four trainee operators performed 15 simulated biopsy procedures on the targets (series one) and repeated identical procedures 2 weeks later (series two). Statistical analysis of accuracy of needle placement and subject confidence were performed. Results: Significant sequential improvement in axial plane angular error was noted with the average error decreasing by 0.33 degrees after every five procedures performed (95% CI: -0.58 to -0.08, p = 0.01). Operator confidence indicated significant improvement both within each series and from series one to series two (95% CI: 0.08 to 1.17, p = 0.025 and 95% CI: 0.00 to 0.58, p = 0.05) respectively. However, variability in operator performance made statistically significant improvement in other variables unproven. Conclusion: Despite the study comprising a relatively small number of participants and procedures, it clearly demonstrated the effectiveness of teaching operators to perform CT-guided procedures using a phantom system. Needle placement accuracy significantly improved, with a reduction in axial angular error, and improved operator confidence without the risks associated with training on patients. Three of the operators in this study had never performed a CT-guided procedure previously, and their proficiency, after a relatively short but intense period of training, was impressive. The use of phantoms should be considered routinely for basic training of CT-guided needle placement

Additional details

Identifiers

DOI
10.1016/j.crad.2006.09.010;
PII
S0009-9260(06)00318-7;

Publishing Information

Journal Title
Clinical Radiology
Journal Volume
62
Journal Issue
2
Journal Page Range
p. 166-171
ISSN
0009-9260
CODEN
CLRAAG

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38089391
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; BIOPSY; COMPUTERIZED TOMOGRAPHY; ERRORS; EVALUATION; PATIENTS; PHANTOMS; TRAINING
Descriptors DEC
DIAGNOSTIC TECHNIQUES; EDUCATION; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY