Published 1995 | Version v1
Journal article

Deterministic analytical method for the calibration of semiconductor detectors in gamma ray spectrometry

  • 1. Centre d'Etude de l'Energie Nucleaire, Mol (Belgium)

Description

A simple to use computer program (called ATT3D) to calculate the detector efficiency for any type of geometry and material is described. The requested data are a point geometry efficiency and the description of the detector's and sample's geometries. This program, adapted for a 486 computer or for a workstation, can be used in different applications and especially in body burden assessments by the direct method (whole-body counting). The accuracy depends on the description of the geometry which can be adapted by changing the mesh size when high precision is requested. (Author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
61
Journal Issue
1-3
Journal Page Range
p. 141-144.
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
applications and research needs in health physics and dosimetry.
Acronym
International workshop on advances in radiation measurements
Dates
3-6 Oct 1994.
Place
Chalk River, ON (Canada).

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
28005824
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
A CODES; CALIBRATION; EFFICIENCY; GAMMA SPECTROSCOPY; GEOMETRY; MATHEMATICAL MODELS; RADIATION PROTECTION; SEMICONDUCTOR DETECTORS; VALIDATION
Descriptors DEC
COMPUTER CODES; MATHEMATICS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SPECTROSCOPY; TESTING

Optional Information

Secondary number(s)
CONF--941027; EUR--16177-EN.