Published May 1, 1999 | Version v1
Journal article

Performance of n-γ pulse-shape discrimination with simple pile-up rejection at high γ-ray count rates

Description

The performance of n-γ pulse-shape discrimination for a liquid scintillation detector has been investigated for γ-ray count rates up to 50 kcps. A method in which the ratio of the total to partial charge in the anode pulse is directly measured has shown much improved quality of the pulse-shape discrimination when pile-up events are rejected; it can discriminate neutron events of 50 cps from γ-ray events of 29 kcps. The method with simple pile-up rejection has the advantage that only general purpose electronics are required

Additional details

Identifiers

PII
S0168900298013825;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
426
Journal Issue
2-3
Journal Page Range
p. 497-502
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 1999 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.