Published 2007 | Version v1
Miscellaneous Open

Alanine-EPR dosimetry system. Why we like it?

Creators

  • 1. Laboratory for Measurement of Technological Doses, Institute of Nuclear Chemistry and Technology, Warsaw (Poland)

Description

To develop a new high-dose dosimeter we should: (1) to find material with radiation effect monotonically (if possible linearly) dependent on an absorbed dose; (2) to investigate its dosimetric characteristics (sensitivity, dose range, repeatability, accuracy, post-effects); (3) to evaluate economical parameters of new method (cost and availability of dosimetric material, cost of analytical instrument and its services); (4) to evaluate operational features of new dosimeter (sensitivity for environmental conditions, time from irradiation to the read-out); (5) to perform a calibration curve, i.e. functional dependence between radiation effect (dosimetric signal) and absorbed dose. On the base of this very stable stable ammonium radical (SAR) generated in crystalline α-alanine was established in the INCT as an alanine-EPR dosimetry system. Presented lecture describes the main features of this dosimeter

Files

39066524.pdf

Files (2.2 MB)

Name Size Download all
md5:9deea94edfb76c1a33d445fb0b60e91b
2.2 MB Preview Download
Part of:
Materials of the Regional Training Course on Validation and Process Control for Electron Beam Radiation Processing

Additional details

Publishing Information

Publisher
Institute of Nuclear Chemistry and Technology
Imprint Place
Warsaw (Poland)
Imprint Title
Materials of the Regional Training Course on Validation and Process Control for Electron Beam Radiation Processing
Imprint Pagination
[557 p.]
Journal Page Range
[36 p.]
Report number
INIS-PL--2008-0008

Conference

Title
Regional Training Course on Validation and Process Control for Electron Beam Radiation Processing
Dates
3-7 Dec 2007
Place
Warsaw (Poland)

Optional Information