Local flow speed measurement using tunable AC thermal anemometry
- 1. Seoul National Univ., Seoul (Korea, Republic of)
- 2. Korea Univ., Seoul (Korea, Republic of)
- 3. Chungang Univ., Seoul (Korea, Republic of)
- 4. Hongik Univ., Seoul (Korea, Republic of)
Description
This paper shows the results of local flow speed measurement using tunable AC thermal anemometry, which is suitable for the accurate measurement of wide range flow speed. The measurement accuracy is verified through the comparison between the measurement data and the analytic solution of the sensor temperature oscillation in stationary fluid. The relation between the phase lag and the flow speed is experimentally investigated at various conditions. The measurement sensitivity for low flow speed improves in a low frequency region and that for high flow speed improves in a high frequency region. Also, the sensitivity increases with decreasing thermal conductivity of the surrounding fluid. The local flow speed could be measured as low as 1.5 mm/s and the highest measurement resolution was 0.05 mm/s in the range of 4.5∼5.0 mm/s at 1 Hz in this experiment
Additional details
Publishing Information
- Journal Title
- Journal of Mechanical Science and Technology
- Journal Volume
- 19
- Journal Issue
- 7
- Series
- 9 refs, 9 figs
- Journal Page Range
- p. 1449-1459
- ISSN
- 1738-494X
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 37056391
- Subject category
- S42: ENGINEERING;
- Descriptors DEI
- ANEMOMETERS; CYLINDERS; FLUIDIC DEVICES; HEAT TRANSFER; LAMINAR FLOW; OSCILLATIONS; SENSITIVITY; THERMAL CONDUCTIVITY
- Descriptors DEC
- ENERGY TRANSFER; FLUID FLOW; MEASURING INSTRUMENTS; PHYSICAL PROPERTIES; THERMODYNAMIC PROPERTIES