Published August 2021 | Version v1
Journal article

A simple spectrofluorimetric method to estimate uranium. Application to nuclear reprocessing solution

  • 1. Materials Chemistry and Metal Fuel Cycle Group, Kalpakkam (India). Analytical Chemistry and Spectroscopy Division
  • 2. Indira Gandhi Centre for Atomic Research, Kalpakkam (India). Homi Bhabha National Institute

Description

A simple spectrofluorimetric method is reported for the estimation of uranium (U) in nitric acid using 0.4 M H2SO4 as an enhancing reagent. The intensity of fluorescence peak of U at 512 nm with excitation at 414 nm shows linearity over 150-6000 µg/mL. Standard addition method is applied to arrive at the estimates of U in simulated dissolver and high active liquid waste solution of nuclear reprocessing plant. The estimate value of U for the sample with high content of other ions is found to be more accurate than the value measured by Davies-Gray method. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
329
Journal Issue
2
Journal Page Range
p. 1061-1066
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
18 refs.