Published June 1998 | Version v1
Journal article

Thermoluminescence dosimetry applied to in vivo dose measurements for total body irradiation techniques

  • 1. Institut de Tecniques Energetiques (INTE), Universitat Politecnica de Catalunya, Barcelona (Spain)
  • 2. Servei de Radiofisica i Radioproteccio, Hospital de la Santa Creu i Sant Pau (HSCSP), Sant Antoni Marie Claret, Barcelona (Spain)

Description

In total body irradiation (TBI) treatments in vivo dosimetry is recommended because it makes it possible to measure the accuracy and quality control of dose delivery. The aim of this work is to set up an in vivo thermoluminescence dosimetry (TLD) system to measure the dose distribution during the TBI technique used prior to bone marrow transplant. Some technical problems due to the presence of lung shielding blocks are discussed. Irradiations were performed in the Hospital de la Santa Creu i Sant Pau by means of a Varian Clinac-1800 linear accelerator with 18 MV X-ray beams. Different TLD calibration experiments were set up to optimize in vivo dose assessment and to analyse the influence on dose measurement of shielding blocks. An algorithm to estimate midplane doses from entrance and exit doses is proposed and the estimated dose in critical organs is compared to internal dose measurements performed in an Alderson anthropomorphic phantom. The predictions of the dose algorithm, even in heterogeneous zones of the body such as the lungs, are in good agreement with the experimental results obtained with and without shielding blocks. The differences between measured and predicted values are in all cases lower than 2%. The TLD system described in this work has been proven to be appropriate for in vivo dosimetry in TBI irradiations. The described calibration experiments point out the difficulty of calibrating an in vivo dosimetry system when lung shielding blocks are used. (author)

Additional details

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
47
Journal Issue
3
Journal Page Range
p. 319-324
ISSN
0167-8140
CODEN
RAONDT

INIS

Country of Publication
Ireland
Country of Input or Organization
Ireland
INIS RN
31044258
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
IN VIVO; RADIOTHERAPY; THERMOLUMINESCENT DOSIMETRY; WHOLE-BODY IRRADIATION
Descriptors DEC
DOSIMETRY; EXTERNAL IRRADIATION; IRRADIATION; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY