Published 1985 | Version v1
Report Open

Method of Moessbauer spectroscopy for 67Zn

  • 1. Bylgarska Akademiya na Naukite, Sofia. Inst. za Yadrena Izsledvaniya i Yadrena Energetika
  • 2. Joint Inst. for Nuclear Research, Dubna (USSR). Lab. of Nuclear Problems

Description

The method of Moessbauer measurement for the 93.3 keV level of 67Zn isotope is developed. For this puspose a high stability digital sinusoidal voltage generator for the 2-900 V range has been constructed. It provides deviation from the sinusoidal law smaller than 3x10-5; frequency and amplitude instability smaller than 10-5 and 10-3, respectively. The frequency response of quartz and piezoelectric transducers was measured and several improvements have been made. A piesoceramic sensor has been mounted at the place of the source-holding spring for permanent movement control. This reveals new possibilities for futher optimising the transducer construction, working regime adjusting and distortion compensation. The accuracy in displacement measurement is 4x10-12 m

Availability note (English)

MF available from INIS under the Report Number.

Files

17055582.pdf

Files (291.4 kB)

Name Size Download all
md5:27e62bfecfaf8a8f536ffc573ff15ed1
291.4 kB Preview Download

System files (38.8 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Методика мессбауэровских измерений на

Publishing Information

Imprint Pagination
10 p.
Report number
JINR--13-85-615

Optional Information

Notes
19 refs.; 5 figs.