Published August 2008 | Version v1
Journal article

Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications

  • 1. Departement de Physique, de Genie Physique et d'Optique, Universite Laval, Quebec, Quebec, G1K 7P4 (Canada) and Departement de Radio-Oncologie, Hotel-Dieu de Quebec, Centre Hospitalier Universitaire de Quebec, Quebec, Quebec G1R2J6 (Canada)
  • 2. Department of Radiation Physics, University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030 (United States)
  • 3. Departement de Physique, de Genie Physique et d'Optique, Universite Laval, Quebec, Quebec G1K 7P4 (Canada) and Departement de Radio-Oncologie, Hotel-Dieu de Quebec, Centre Hospitalier Universitaire de Quebec, Quebec, Quebec G1R2J6 (Canada)

Description

A clinical prototype of a scintillating fiber dosimeter array for quality assurance applications is presented. The array consists of a linear array of 29 plastic scintillation detectors embedded in a water-equivalent plastic sheet coupled to optical fibers used to guide optical photons to a charge coupled device (CCD) camera. The CCD is packaged in a light-tight, radiation-shielded housing designed for convenient transport. A custom designed connector is used to ensure reproducible mechanical positioning of the optical fibers relative to the CCD. Profile and depth dose characterization measurements are presented and show that the prototype provides excellent dose measurement reproducibility (±0.8%) in-field and good accuracy (±1.6% maximum deviation) relative to the dose measured with an IC10 ionization chamber

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
35
Journal Issue
8
Journal Page Range
p. 3682-3690
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2008 American Institute of Physics