Electron arc therapy: beam data requirements and treatment planning
- 1. Johns Hopkins Univ., Baltimore, MD (USA)
Description
In electron arc therapy a long narrow beam defined by a secondary collimator is used to sweep across the treatment area of the patient. Central axis percentage depth doses and off-axis ratios measured with such a collimator show variations with the source-to-surface distance (SSD). Errors may be introduced into isodose distributions calculated with the effective SSD method when beam data measured at a fixed SSD are used. A treatment planning algorithm using beam data measured at two SSDs to allow an interpolation and extrapolation procedure to account for SSD variations has been implemented. For calculations in off-axis planes, off-axis ratios normal to the plane of rotation are also required. Results of the calculations were checked with film dosimetry measurements and found to be in good agreement. (author)
Additional details
Publishing Information
- Journal Title
- Clin. Radiol.
- Journal Volume
- 38
- Journal Issue
- 4
- Series
- Clin. Radiol.
- Journal Page Range
- 379-383
- ISSN
- 0009-9260
- CODEN
- CLRAA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 19001807
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; BEAM OPTICS; COLLIMATORS; ELECTRON BEAMS; ERRORS; ISODOSE CURVES; PLANNING; RADIOTHERAPY; SPATIAL DOSE DISTRIBUTIONS
- Descriptors DEC
- BEAMS; DISTRIBUTION; LEPTON BEAMS; MEDICINE; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; THERAPY