Electron arc therapy: beam data requirements and treatment planning

  • 1. Johns Hopkins Univ., Baltimore, MD (USA)

Description

In electron arc therapy a long narrow beam defined by a secondary collimator is used to sweep across the treatment area of the patient. Central axis percentage depth doses and off-axis ratios measured with such a collimator show variations with the source-to-surface distance (SSD). Errors may be introduced into isodose distributions calculated with the effective SSD method when beam data measured at a fixed SSD are used. A treatment planning algorithm using beam data measured at two SSDs to allow an interpolation and extrapolation procedure to account for SSD variations has been implemented. For calculations in off-axis planes, off-axis ratios normal to the plane of rotation are also required. Results of the calculations were checked with film dosimetry measurements and found to be in good agreement. (author)

Additional details

Publishing Information

Journal Title
Clin. Radiol.
Journal Volume
38
Journal Issue
4
Series
Clin. Radiol.
Journal Page Range
379-383
ISSN
0009-9260
CODEN
CLRAA

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
19001807
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ALGORITHMS; BEAM OPTICS; COLLIMATORS; ELECTRON BEAMS; ERRORS; ISODOSE CURVES; PLANNING; RADIOTHERAPY; SPATIAL DOSE DISTRIBUTIONS
Descriptors DEC
BEAMS; DISTRIBUTION; LEPTON BEAMS; MEDICINE; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; THERAPY