Published January 1977 | Version v1
Journal article

Neutron spectrometry by sandwich surface barrier technique

Creators

  • 1. Bhabha Atomic Research Centre, Bombay

Description

A neutron spectrometer using silicon surface barrier detectors employed 6LiF as a radiating material is developed. An incident neutron interacts with the radiating atom resulting in two charged particles, which are detected by sandwich detector geometry and the corresponding sum pulses from both the detectors shall uniquely define the incident neutron energy. Results obtained by Monte Carlo simulation of all the correction factors for the detector head employed are presented. A new concept for resolution correction is developed which brings down the lower energy limit of the detector system. The spectrometer is employed for incore spectrum studies in case of a thermal reactor ''Kahter'' and the results obtained exhibit a good agreement with that of calculations by GAM THERMOS EXTERMINATOR Code

Additional details

Additional titles

Augmented title (English)
"6LiF radiator and Si surface barrier detectors

Publishing Information

Journal Title
Natl. Bur. Stand. (U.S.), Spec. Publ.
Journal Issue
no.461
Series
Natl. Bur. Stand. (U.S.), Spec. Publ.

Conference

Title
International symposium on radiation physics.
Dates
30 Nov 1974.
Place
Calcutta, India.

Optional Information

Notes
See CONF-741109--.