Deformation T-Cup: A new multi-anvil apparatus for controlled strain-rate deformation experiments at pressures above 18 GPa
Creators
- 1. Department of Earth Sciences, University College London, Gower Street, London WC1E 6BT (United Kingdom)
- 2. Mineral Physics Institute, Stony Brook University, Stony Brook, New York 11794-2100 (United States)
Description
A new multi-anvil deformation apparatus, based on the widely used 6-8 split-cylinder, geometry, has been developed which is capable of deformation experiments at pressures in excess of 18 GPa at room temperature. In 6-8 (Kawai-type) devices eight cubic anvils are used to compress the sample assembly. In our new apparatus two of the eight cubes which sit along the split-cylinder axis have been replaced by hexagonal cross section anvils. Combining these anvils hexagonal-anvils with secondary differential actuators incorporated into the load frame, for the first time, enables the 6-8 multi-anvil apparatus to be used for controlled strain-rate deformation experiments to high strains. Testing of the design, both with and without synchrotron-X-rays, has demonstrated the Deformation T-Cup (DT-Cup) is capable of deforming 1–2 mm long samples to over 55% strain at high temperatures and pressures. To date the apparatus has been calibrated to, and deformed at, 18.8 GPa and deformation experiments performed in conjunction with synchrotron X-rays at confining pressures up to 10 GPa at 800 °C
Additional details
Identifiers
- DOI
- 10.1063/1.4891338;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 85
- Journal Issue
- 8
- Journal Page Range
- p. 085103-085103.10
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46020866
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACTUATORS; CALIBRATION; DEFORMATION; PRESSURE RANGE GIGA PA; PRESSURE RANGE MEGA PA 10-100; STRAIN RATE; SYNCHROTRONS; TEMPERATURE RANGE 0273-0400 K; TEMPERATURE RANGE 0400-1000 K; TESTING; X RADIATION
- Descriptors DEC
- ACCELERATORS; CYCLIC ACCELERATORS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; PRESSURE RANGE; PRESSURE RANGE MEGA PA; RADIATIONS; TEMPERATURE RANGE
Optional Information
- Notes
- (c) 2014 Author(s)