Published February 2010 | Version v1
Journal article

Radiometric comparison for measuring the absolute radiant power of a free-electron laser in the extreme ultraviolet

  • 1. National Institute of Advanced Industrial Science and Technology (AIST), NMIJ, Tsukuba (Japan)
  • 2. RIKEN, XFEL Project Head Office, Kouto, Sayo, Hyogo (Japan)
  • 3. Deutsches Elektronen-Synchrotron, DESY, Hamburg (Germany)
  • 4. Physikalisch-Technische Bundesanstalt, PTB, Berlin (Germany)
  • 5. Ioffe Physico-Technical Institute, St Petersburg (Russian Federation)
  • 6. Japan Synchrotron Radiation Research Institute, Sayo, Hyogo (Japan)

Description

At the new SPring-8 extreme ultraviolet free-electron laser in Japan, a radiometric comparison between a gas-monitor detector from Physikalisch-Technische Bundesanstalt/Deutsches Elektronen-Synchrotron/Ioffe Physico-Technical Institute (Ioffe) and a cryogenic radiometer from National Institute of Advanced Industrial Science and Technology was carried out. Radiant power in the range between 30 μW and 350 μW was measured at the wavelengths of 51.3 nm, 56.1 nm and 61.2 nm. The results obtained with the two different detectors agree within 2.6%, which is well below their combined relative standard uncertainty, which varies from 4.1% to 5.5%. (authors)

Availability note (English)

Available from doi: <http://dx.doi.org/10.1088/0026-1394/47/1/003

Additional details

Identifiers

Publishing Information

Journal Title
Metrologia
Journal Volume
47
Journal Issue
no.1
Journal Page Range
p. 21-23
ISSN
0026-1394
CODEN
MTRGAU

Optional Information

Notes
13 refs.