Published February 2010
| Version v1
Journal article
Radiometric comparison for measuring the absolute radiant power of a free-electron laser in the extreme ultraviolet
Creators
- Norio, Saito1
- Masahiro, Kato1
- Norio, Saito2
- Juranic, P.N.2
- Masahiro, Kato2
- Richter, M.2
- Sorokin, A.A.2
- Kai, Tiedtke2
- Mitsuru, Nagasono2
- Makina, Yabashi2
- Kensuke, Tono2
- Tadashi, Togashi2
- Tetsuya, Ishikawa2
- Juranic, P.N.3
- Sorokin, A.A.3
- Kai, Tiedtke3
- Jastrow, U.3
- Richter, M.4
- Sorokin, A.A.4
- Kroth, U.4
- Schoppe, H.4
- Sorokin, A.A.5
- Tadashi, Togashi6
- Hiroaki, Kimura6
- Haruhiko, Ohashi6
- 1. National Institute of Advanced Industrial Science and Technology (AIST), NMIJ, Tsukuba (Japan)
- 2. RIKEN, XFEL Project Head Office, Kouto, Sayo, Hyogo (Japan)
- 3. Deutsches Elektronen-Synchrotron, DESY, Hamburg (Germany)
- 4. Physikalisch-Technische Bundesanstalt, PTB, Berlin (Germany)
- 5. Ioffe Physico-Technical Institute, St Petersburg (Russian Federation)
- 6. Japan Synchrotron Radiation Research Institute, Sayo, Hyogo (Japan)
Description
At the new SPring-8 extreme ultraviolet free-electron laser in Japan, a radiometric comparison between a gas-monitor detector from Physikalisch-Technische Bundesanstalt/Deutsches Elektronen-Synchrotron/Ioffe Physico-Technical Institute (Ioffe) and a cryogenic radiometer from National Institute of Advanced Industrial Science and Technology was carried out. Radiant power in the range between 30 μW and 350 μW was measured at the wavelengths of 51.3 nm, 56.1 nm and 61.2 nm. The results obtained with the two different detectors agree within 2.6%, which is well below their combined relative standard uncertainty, which varies from 4.1% to 5.5%. (authors)
Availability note (English)
Available from doi: <http://dx.doi.org/10.1088/0026-1394/47/1/003Additional details
Identifiers
Publishing Information
- Journal Title
- Metrologia
- Journal Volume
- 47
- Journal Issue
- no.1
- Journal Page Range
- p. 21-23
- ISSN
- 0026-1394
- CODEN
- MTRGAU
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 41110010
- Subject category
- S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- CALIBRATION STANDARDS; DATA COVARIANCES; EXTREME ULTRAVIOLET RADIATION; FREE ELECTRON LASERS; PHOTOIONIZATION; RADIOMETERS; SPRING-8 STORAGE RING; SYNCHROTRON RADIATION
- Descriptors DEC
- BREMSSTRAHLUNG; ELECTROMAGNETIC RADIATION; IONIZATION; LASERS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATION SOURCES; RADIATIONS; STANDARDS; STORAGE RINGS; SYNCHROTRON RADIATION SOURCES; ULTRAVIOLET RADIATION
Optional Information
- Notes
- 13 refs.