Sensitive Mass Spectrometer for Time-Resolved Gas-Phase Chemistry Studies at High Pressures
Creators
- 1. Sandia National Laboratory (SNL-CA), Livermore, CA (United States)
Description
Here, we report the construction of a new experimental apparatus for direct time-resolved probing of high-pressure gas-phase chemical reactions by photoionization mass spectrometry. The apparatus uses a laser photolysis slow-flow reactor, capable of operating at P = 0.3 — 100 bar and T = 300 — 1000 K. In this report, we initiate reactions in homogeneous gas mixtures by the photolysis of appropriate radical precursor using laser pulses at repetition rates of 1 — 10 Hz. The reacting mixture is continuously sampled into a vacuum chamber, ionized by VUV photons from laboratory-based discharge lamps or from a synchrotron beamline, and analyzed by a custom-designed mass spectrometer. Soft near-threshold ionization by tunable synchrotron radiation enables spectroscopic quantification of many key intermediates and products of chemical reactions. A novel ionization scheme in the high-density region of the sample gas jet increases the experimental sensitivity 100-fold, compared with existing instruments, without compromising mass resolution. A 40-kHz pulsed reflectron time-of-flight mass spectrometer in the orthogonal acceleration geometry achieves simultaneous detection of all ionized species with 25-μs time resolution. We show the power of this apparatus by investigating the ethyl radical oxidation reaction using very dilute (<1012 molecules • cm-3) ethyl concentrations at pressures up to 25 bar.
Availability note (English)
Available from https://www.osti.gov/servlets/purl/1574695; https://www.osti.gov/biblio/1574695; DOE Accepted Manuscript full text, or the publishers Best Available Version will be available free of charge after the embargo periodAdditional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physical Chemistry. A, Molecules, Spectroscopy, Kinetics, Environment, and General Theory
- Journal Volume
- 123
- Journal Issue
- 50
- Journal Page Range
- p. 10804-10814
- ISSN
- 1089-5639
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 53046626
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S47: OTHER INSTRUMENTATION;
- Descriptors DEI
- ETHYL RADICALS; FAR ULTRAVIOLET RADIATION; HOMOGENEOUS MIXTURES; MASS RESOLUTION; MASS SPECTROSCOPY; OXIDATION; PHOTOIONIZATION; PHOTOLYSIS; SENSITIVITY; SYNCHROTRON RADIATION; TIME RESOLUTION; TIME-OF-FLIGHT MASS SPECTROMETERS
- Descriptors DEC
- ALKYL RADICALS; BREMSSTRAHLUNG; CHEMICAL REACTIONS; DECOMPOSITION; DISPERSIONS; DYNAMIC MASS SPECTROMETERS; ELECTROMAGNETIC RADIATION; IONIZATION; MASS SPECTROMETERS; MEASURING INSTRUMENTS; MIXTURES; PHOTOCHEMICAL REACTIONS; RADIATIONS; RADICALS; RESOLUTION; SPECTROMETERS; SPECTROSCOPY; TIME-OF-FLIGHT SPECTROMETERS; TIMING PROPERTIES; ULTRAVIOLET RADIATION
Optional Information
- Contract/Grant/Project number
- AC04-94AL85000; NA0003525; AC02-05CH11231
- Funding organization
- USDOE Office of Science - SC, Basic Energy Sciences (BES). Chemical Sciences, Geosciences, and Biosciences Division (United States)
- Secondary number(s)
- OSTIID--1574695