Published May 15, 1976
| Version v1
Journal article
A design of a velocimeter using a radioactive positron emitting source without lead collimators
Description
A possible method of eliminating lead collimators without loss of spatial accuracy in the velocity or position measurement of a positron active source is described here, and the accuracies of this and the lead collimator method are calculated and compared. The theoretical calculations have been tested by laboratory measurements, It is shown that at low source activities the method described is superior to the conventional lead collimator method. (Auth.)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Instruments and Methods
- Journal Volume
- 135
- Journal Issue
- 1
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 61-65
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 7269234
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANNIHILATION; GAMMA DETECTION; GAMMA RADIATION; MEASURING METHODS; POSITRON SOURCES; POSITRONS; VELOCITY
- Descriptors DEC
- ANTILEPTONS; ANTIMATTER; ANTIPARTICLES; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; INTERACTIONS; IONIZING RADIATIONS; LEPTONS; MATTER; PARTICLE INTERACTIONS; PARTICLE SOURCES; RADIATION DETECTION; RADIATION SOURCES; RADIATIONS
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent