Description

A Siemens electronic portal imaging system called BeamviewplusTM was used daily in 15 lung cancer patients for each X-ray field. Out of the 714 expected portal images, 585 (82%), were obtained and printed. We counted 94 errors in the 585 successful controls (16 %). Four main classes of errors were identified: irradiation field placement (47 cases, 50 %), shielding placement (30 cases, 32 %), patients's position on the tilt examining table (14 cases, 15 %), and selection of the energy (3 cases, (3 %). Forty percent of these errors were potentially serious in terms of treatment volume and vital organ protection. Only portal imaging control could easily detect errors because they occurred randomly. We thus consider the use of this daily system helpful for patient management. (author)

Additional details

Additional titles

Original title (French)
Utilisation d'un systeme d'imagerie en temps reel dans le controle quotidien de patients traites par irradiation pour un cancer thoracique

Publishing Information

Journal Title
Cancer Radiotherapie
Journal Volume
1
Journal Issue
no.3
Series
30 refs.
Journal Page Range
p. 249-257
ISSN
1278-3218
CODEN
CARAFC

INIS

Country of Publication
France
Country of Input or Organization
France
INIS RN
29041851
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CARCINOMAS; CHEST; IMAGES; NUCLEAR MEDICINE; PORTAL SYSTEM; QUALITY ASSURANCE; RADIOTHERAPY
Descriptors DEC
BLOOD VESSELS; BODY; BODY AREAS; CARDIOVASCULAR SYSTEM; DISEASES; MEDICINE; NEOPLASMS; ORGANS; THERAPY; VEINS