Performance of the readout system for MONOLITH

Description

In this paper, we describe the performance of the readout system for MONOLITH developed at the LNGS. This system is based on the use of flat cables as readout elements, instead of the conventional copper strips. The advantages of flat cable strips are the good performance, the easy installation and the possibility to realize complex readout systems. The X-coordinate readout system (X-system) is composed by 15 m long, Flat Cable Strips (FCS). The distribution of the time difference between the streamer signals transmitted at both the ends of the X-system FCS has a sigma resolution of the order of 100 ps. This resolution allows the measurement of the particle direction by means of the time-of-flight technique and can be exploited to measure the Y-coordinate with a resolution in the order of 1 cm. The Y-coordinate system is composed by short FCS connected together by a flat cable acting as a bus line. It allows the installation of the electronics outside the apparatus minimizing the number of channels

Additional details

Identifiers

PII
S0168900203013500;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
508
Journal Issue
1-2
Journal Page Range
p. 194-198
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
6. international workshop on resistive plate chambers and related detectors
Dates
26-27 Nov 2001
Place
Coimbra (Portugal)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35021564
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COORDINATES; PARTICLE DISCRIMINATION; READOUT SYSTEMS; RESOLUTION; SIGNAL CONDITIONING; TIME-OF-FLIGHT METHOD
Descriptors DEC
PARTICLE IDENTIFICATION

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.