Published June 1981 | Version v1
Journal article

Infrared measurement of the temperature distribution of a repetitively pulsed electron beam extraction foil as a beam uniformity diagnostic

  • 1. Sandia Natl Lab, Albuquerque, NM

Description

A presentation is made of experiments to measure the current extracted through the foil of an electron beam diode by observing the heating of the foil. Generally anode current density is measured with a dosimeter film, often of dubious linearity, or with Faraday cups which give a current density versus time map at isolated points or with an array of calorimeters. For the present measurements a fast scanning infrared camera is used to measure the temperature rise of the foil after a pulse. The measurement takes 600/mu/s for a single line scan or 16 ms for a full raster scan. 5 refs

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-28
Journal Issue
3
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
3392-3394
ISSN
0018-9499

Conference

Title
Particle accelerator conference.
Dates
11 - 13 Mar 1981.
Place
Washington, DC (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14724782
Subject category
S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BEAM EXTRACTION; BEAM MONITORING; ELECTRON BEAMS; FOILS; INFRARED THERMOGRAPHY; PULSE TECHNIQUES; TEMPERATURE MEASUREMENT
Descriptors DEC
BEAMS; LEPTON BEAMS; MONITORING; PARTICLE BEAMS

Optional Information

Secondary number(s)
CONF-810314--.