Published January 1992 | Version v1
Journal article

Quantitative analysis in single photon emission tomography (SPET)

  • 1. Rijksuniversiteit Leiden (Netherlands). Div. of Nuclear Medicine
  • 2. Rijksuniversiteit Leiden (Netherlands). Lab. for Clinical and Experimental Image Processing

Description

Quantitative analysis can improve the sensitivity and specificity of single photon emission tomography (SPET) procedures, as well as reduce inter- and intraobserver variabilities. Quantification of the radioactivity distribution is the ultimate goal of SPET. In this review we consider the basic requirements for an optimum three-dimensional reconstruction of the radionuclide distribution to enable quantification. Attenuation and scatter correction as well as varying resolution are the major problems. In the older SPET systems quantification was hampered by the lack of system sensitivity and sufficient computer power. Therefore, the imaging system was often assumed to the shift invariant and linear and the attenuation throughout the object uniform. More sophisticated solutions have been proposed and with more or less success implemented, but not for application in daily practice. Knowledge (measurement) of the attenuation is often required. New generation SPET systems employing multi-detectors and super minicomputers will ease the implementation of these solutions. (orig.)

Additional details

Publishing Information

Journal Title
European Journal of Nuclear Medicine
Journal Volume
19
Journal Issue
1
Series
Eur. J. Nucl. Med.
Journal Page Range
47-61
ISSN
0340-6997
CODEN
EJNMD