Published March 2000 | Version v1
Journal article

Comparison of scatter and attenuation correction methods in single photon emission CT. Simulation study

  • 1. Hosei Univ., Koganei, Tokyo (Japan). Coll. of Engineering

Description

Quantitative SPECT requires an accurate and appropriate attenuation correction and scatter correction. This paper compares scatter corrected images combined with attenuation correction. The scatter correction methods used here were the deconvolution method, dual energy window subtraction method, triple energy window method, and dual photopeak window method. The attenuation correction methods used were Sorenson's method, iterative Chang's method, and Bellini's method. The numerical phantoms used were the cold spot, hot spot, star, and MCAT phantom. These correction methods were assessed by using profiles and images, contrast values, and linearity of SPECT values. (author)

Additional details

Publishing Information

Journal Title
Kaku Igaku
Journal Volume
37
Journal Issue
2
Journal Page Range
p. 131-142
ISSN
0022-7854