Published September 1986 | Version v1
Journal article

Rectangular waveguide calorimeter for single intense microwave pulses

  • 1. Microwave Physics Division, Sandia National Laboratories, Albuquerque, New Mexico 87185

Description

A new calorimeter for single intense microwave pulses has been designed and tested. The device was constructed in WR284 rectangular waveguide and was operated at frequencies from 2.5 to 4.3 GHz with a tunable instantaneous bandwidth of approximately 500 MHz. The calorimeter used a single thermistor to measure the energy deposited on a carbon absorber having a microwave power absorbance of 90%. The calorimeter was tested at power levels from 100 to 1000 MW for pulse lengths of 12--6 ns, respectively. The sensitivity of the device was 200--300 mV/J

Additional details

Publishing Information

Journal Title
Rev. Sci. Instrum.
Journal Volume
57
Journal Issue
9
Series
Rev. Sci. Instrum.
Journal Page Range
2359-2361
ISSN
0034-6748
CODEN
RSINA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17089969
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALORIMETERS; CALORIMETRY; MICROWAVE RADIATION; OPERATION; POWER; SPECIFICATIONS; TESTING; WAVEGUIDES
Descriptors DEC
ELECTROMAGNETIC RADIATION; MEASURING INSTRUMENTS; RADIATIONS