Published 1967 | Version v1
Report Open

Numerical counting ratemeter with large time constant for very low count rates - operational results

  • 1. Commissariat a l'Energie Atomique, Saclay (France). Centre d'Etudes Nucleaires

Description

The use of an analogue ictometer poses technical problems for measuring low count rates. These latter call for time constants of up to 1000 sec. The numerical technique is better adapted and makes it possible to increase the time constant to 1000 sec. or over. The equipment described consists of 6 numerical ictometers fitted in a 3 U 3 drawer. By means of a clock it is possible to vary the counting time from 1 second to 1000 seconds, using values at 5, 10, 50, 100 and 500 seconds. The corresponding values of the frequency are 500 c/sec, 100, 50 10. 5. 1 and 0.5 c/sec with a resolving power of 0.2 per cent. The result appears on an indicator panel (galvanometer) and is available in the form of an analogue voltage suitable for a 50 mV MECI recorder. (author)

Availability note (English)

Available from INIS in electronic form

Abstract (French)

En utilisant un ictometre analogique, la mesure des taux de comptages faibles pose des problemes techniques. Les faibles taux de comptage exigent des constantes de temps allant jusqu'a 1000 s. La technique numerique se montre mieux adaptee et permet d'augmenter la constante de temps a 1000 s ou plus. L'ensemble decrit comporte 6 ictometres numeriques loges dans un tiroir 3 U 3. Une horloge permet de varier le temps de comptage de 1 seconde a 1000 secondes, passant par 5, 10, 50, 100, 500 secondes. Les gammes de frequences correspondantes sont 500, 100, 50, 10, 5, 1, 0,5 c/s avec une resolution de 2 pour 1000. Le resultat est affiche par un indicateur de tableaux (galvanometre) et est disponible en tension analogique pour l'enregistrement sur MECI (50 mV). (auteur)

Files

37038421.pdf

Files (388.8 kB)

Name Size Download all
md5:1c054e3cb4f0bc6e1318630175b84969
388.8 kB Preview Download

Additional details

Additional titles

Original title (French)
Ictometre numerique a grande constante de temps pour tres faible taux de comptage - resultats d'exploitation

Publishing Information

Imprint Pagination
16 p.
Report number
CEA-R--3340

INIS

Country of Publication
France
Country of Input or Organization
France
INIS RN
37038421
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CORRECTIONS; COUNTING RATEMETERS; COUNTING RATES; DIGITAL SYSTEMS; MEASURING INSTRUMENTS; NEUTRON FLUX; TIME DEPENDENCE
Descriptors DEC
ELECTRONIC EQUIPMENT; EQUIPMENT; RADIATION FLUX

Optional Information

Notes
0 refs.