Published August 2003 | Version v1
Journal article

Measurements of free radicals over a period of 4.5 years in gamma-irradiated ultra-high molecular weight polyethylene

Description

Gamma irradiation-induced free radicals have been investigated for 4.5 years in medical grade ultra-high molecular weight polyethylene resins, GUR 4120, GUR 4150 and Himont 1900, at room temperature (23 deg. C) in absence (vacuum) or presence of oxygen (air), and at 75 deg. C in vacuum. In vacuum, the primary radicals (predominantly allyl) decayed following Y=YR+ae-bt and, in 30 days at 75 deg. C or 100 days at 23 deg. C, their concentration was reduced from ∼1017 radical/g (initial) to a minimum YR∼1015 radical/g and remained at this level as long as the samples were kept in vacuum. When the samples were exposed to air, at any time during this study, these residual radicals were found to decay to a very stable, polyenyle-type terminal radical

Additional details

Identifiers

PII
S0168583X03006451;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
Journal Volume
208
Journal Issue
3
Journal Page Range
p. 204-209
ISSN
0168-583X
CODEN
NIMBEU

Conference

Title
5. conference on ionizing radiation and polymers
Dates
21-26 Sep 2002
Place
Sainte-Adele, PQ (Canada)

INIS

Country of Publication
Netherlands
Country of Input or Organization
Pakistan
INIS RN
35054250
Subject category
S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GAMMA RADIATION; MOLECULAR WEIGHT; OXIDATION; POLYETHYLENES; POLYPHENYLS; RADICALS; RADIOSTERILIZATION
Descriptors DEC
AROMATICS; CHEMICAL REACTIONS; ELECTROMAGNETIC RADIATION; HYDROCARBONS; IONIZING RADIATIONS; IRRADIATION; ORGANIC COMPOUNDS; ORGANIC POLYMERS; POLYMERS; POLYOLEFINS; RADIATIONS; STERILIZATION

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.