Published January 1995 | Version v1
Journal article

Performance of the upgraded multichannel neutron collimator

  • 1. Princeton University, Plasma Physics Laboratory, Princeton, New Jersey 08543 (United States)

Description

The original ZnS based NE451 detectors become saturated at the highest levels of D-D operation. A new less sensitive detector configuration, also based on ZnS crystals, was designed and tested at PPPL in order to extend the collimator capabilities to D-T plasma operation. The new detector design is composed of a thin layer of ZnS powder uniformly deposited between two layers of Plexiglas. The detector capabilities are further extended by upgraded signal processing electronics for a total enhancement of ∼1000x compared to the original detector system. An array of these detectors has been installed on TFTR in tandem with the original detector system. The current mode is presently being implemented on the original NE451 detectors. This will extend their capabilities and allow for greater overlap in data taken in the crossover region between the two detector systems. The performance of the collimator system will be illustrated by neutron data obtained during D-T operation

Additional details

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
66
Journal Issue
1
Journal Page Range
p. 916-918.
ISSN
0034-6748
CODEN
RSINAK

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
26035549
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Descriptors DEI
COLLIMATORS; NEUTRON DETECTORS; PERFORMANCE; PLASMA DIAGNOSTICS; SIGNAL CONDITIONING; SIMULATION; TFTR TOKAMAK
Descriptors DEC
CLOSED PLASMA DEVICES; MEASURING INSTRUMENTS; RADIATION DETECTORS; THERMONUCLEAR DEVICES; TOKAMAK DEVICES