A Virtual Frame System for Stereotactic Radiosurgery Planning
Creators
- 1. Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD (United States)
- 2. Department of Neurosurgery, Johns Hopkins University, Baltimore, MD (United States)
Description
Purpose: We describe a computerized (or virtual) model of a stereotactic head frame to enable planning prior to the day of radiosurgery. The location of the virtual frame acts as a guide to frame placement on the day of the procedure. Methods and Materials: The software consists of a triangular mesh representation of the essential frame hardware that can be overlaid with any MR scan of the patient and manipulated in three dimensions. The software calculates regions of the head that will actually be accessible for treatment, subject to the geometric constraints of the Leksell Gamma Knife hardware. DICOM-compliant MR images with virtual fiducial markers overlaid onto the image can then be generated for recognition by the treatment planning system. Results: Retrospective evaluation of the software on 24 previously treated patients shows a mean deviation of the position of the virtual frame from the actual frame position of 1.6 ± 1.3 mm. Initial clinical use on five patients indicates an average discrepancy of the virtual frame location and the actual frame location of <1 mm. MR images with virtual fiducial markers can be imported into radiosurgical treatment planning software and used to generate an initial treatment plan. Conclusions: The virtual frame provides a tool for prospective determination of lesion accessibility, optimization of the frame placement, and treatment planning before the day of the procedure. This promises to shorten overall treatment times, improve patient comfort, and reduce the need for repeat treatments due to suboptimally placed frames
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ijrobp.2008.06.1934Additional details
Identifiers
- DOI
- 10.1016/j.ijrobp.2008.06.1934;
- PII
- S0360-3016(08)03058-7;
Publishing Information
- Journal Title
- International Journal of Radiation Oncology, Biology and Physics
- Journal Volume
- 72
- Journal Issue
- 4
- Journal Page Range
- p. 1244-1249
- ISSN
- 0360-3016
- CODEN
- IOBPD3
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 40047660
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTER CODES; EVALUATION; HEAD; IMAGES; OPTIMIZATION; PATIENTS; PLANNING; RADIOTHERAPY; SURGERY
- Descriptors DEC
- BODY; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY