Published 1996
| Version v1
Journal article
Studies on the practical application of electret ionisation chambers for radon dosimetry under special conditions
- 1. Technische Univ. Dresden (Germany)
- 2. Research Institute of Preventive and Clinical Medicine, Bratislava (Slovakia)
Description
Results of experimental and theoretical studies are presented analysing the usability of E-PERM electret ionisation chambers under special conditions. High humidity does not influence the readings of E-PERM detectors if additional stress to the electrets, such as rapid temperature changes, is avoided and the readout takes place under normal conditions. The application of the detectors at higher altitudes requires the use of correction factors. The calculated altitude correction factors depend on the volume of the chambers. The behaviour of the electret ionisation chambers during short-term exposures is discussed and appropriate correction factors are given. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 63
- Journal Issue
- 4
- Journal Page Range
- p. 275-279.
- ISSN
- 0144-8420
- CODEN
- RPDODE
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 27057012
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CORRECTIONS; DOSEMETERS; DOSIMETRY; ELECTRETS; EMANOMETERS; HUMIDITY; IONIZATION CHAMBERS; LEVELS; MATHEMATICAL MODELS; PERFORMANCE; RADON
- Descriptors DEC
- DIELECTRIC MATERIALS; ELEMENTS; MATERIALS; MEASURING INSTRUMENTS; MOISTURE; NONMETALS; RADIATION DETECTORS; RARE GASES