Published February 1981
| Version v1
Journal article
The relationship between transmission efficiencies in the FRR and FAT modes of an electron spectrometer
Creators
- 1. Imperial Chemical Industries Ltd., Runcorn (UK). Corporate Lab.
- 2. Surrey Univ., Guildford (UK)
Description
The electron transmission efficiency as a function of kinetic energy for an electron spectrometer operating in the fixed analyser transmission (FAT) mode has been determined by direct comparison with the known transmission efficiency in the fixed retard ratio (FRR) mode. Results obtained from the background intensity and from the relative peak heights of a number of elements compare well with each other. They suggest that transmission in the FAT mode is nearly independent of kinetic energy. (orig.)
Additional details
Publishing Information
- Journal Title
- J. Electron Spectrosc. Relat. Phenom.
- Journal Volume
- 22
- Journal Issue
- 1
- Series
- J. Electron Spectrosc. Relat. Phenom.
- Journal Page Range
- 53-60
- ISSN
- 0368-2048
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 13651084
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- EFFICIENCY; ELECTRON SPECTROMETERS; ELECTRONS; ENERGY DEPENDENCE; KINETIC ENERGY; PHOTOELECTRON SPECTROSCOPY; TRANSMISSION
- Descriptors DEC
- ELECTRON SPECTROSCOPY; ELEMENTARY PARTICLES; ENERGY; FERMIONS; LEPTONS; MEASURING INSTRUMENTS; SPECTROMETERS; SPECTROSCOPY