Investigation of the optimal backscatter for an aSi electronic portal imaging device
Creators
- 1. Department of Radiation Oncology, Medical College of Virginia Hospitals, Virginia Commonwealth University, Richmond, VA 23298 (United States)
Description
The effects of backscattered radiation on the dosimetric response of the Varian aS500 amorphous silicon electronic portal imaging device (EPID) are studied. Measurements demonstrate that radiation backscattered from the EPID mechanical support structure causes 5% asymmetries in the detected signal. To minimize the effect of backscattered radiation from the support structure, this work proposes adding material downstream of the EPID phosphor which provides uniform backscattering material to the phosphor and attenuates backscatter from the support structure before it reaches the phosphor. Two material locations were studied: downstream of the existing image cassette and within the cassette, immediately downstream of the flat-panel imager glass panel. Monte Carlo simulations were used to determine the thicknesses of water, Pb and Cu backscattering materials required to saturate the backscattered signal response for 6 MV and 18 MV beams for material thicknesses up to 50 mm. Water was unable to saturate the backscattered signal for thicknesses up to 50 mm for both energies. For Pb, to obtain a signal within 1% of saturation, 3 mm was required at 6 MV, and 6.8 mm was required at 18 MV. For Cu, thicknesses of 20.6 mm and 22.6 mm were required for the 6 MV and 18 MV beams, respectively. For saturation thicknesses, at 6 MV, the Cu backscatter enhanced the signal more than for Pb (Cu 1.25, Pb 1.11), but at 18 MV the reverse was found (Cu 1.19, Pb 1.23). This is due to the fact that at 6 MV, the backscattered radiation signal is dominated by low-energy scattered photons, which are readily attenuated by the Pb, while at 18 MV, electron backscatter contributes substantially to the signal. Image blurring caused by backscatter spread was less for Pb than Cu. Placing Pb immediately downstream of the glass panel further reduced the signal spread and increased the backscatter enhancement to 1.20 and 1.39 for the 6 MV and 18 MV beams, respectively. Overall, it is determined that adding ∼5 mm of Pb between the detector and the mechanical support structure will substantially reduce the nonuniformity in the backscattered signals for 6 MV and 18 MV photon beams
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/49/1723/pmb4_9_010.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/49/1723/pmb4_9_010.pdf; http://www.iop.org/;
- DOI
- 10.1088/0031-9155/49/9/010;
- PII
- S0031-9155(04)72470-7;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 49
- Journal Issue
- 9
- Journal Page Range
- p. 1723-1738
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35070647
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ASYMMETRY; BACKSCATTERING; BIOMEDICAL RADIOGRAPHY; COMPUTERIZED SIMULATION; IMAGES; MONTE CARLO METHOD; PHOTON BEAMS; PORTAL SYSTEM; SILICON; THICKNESS; WATER
- Descriptors DEC
- BEAMS; BLOOD VESSELS; BODY; CALCULATION METHODS; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DIMENSIONS; ELEMENTS; HYDROGEN COMPOUNDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; OXYGEN COMPOUNDS; RADIOLOGY; SCATTERING; SEMIMETALS; SIMULATION; VEINS