Deflections in computed dose distribution of I-125 seed implant due to angularly averaged approximation
Creators
- 1. Helsinki Univ. Central Hospital (Finland). Dept. of Obstetrics and Gynecology
Description
Dose rate distribution of commercially available I-125 seeds is strongly anisotropic due to self absorption. Effects of this anisotropy to dose distribution of linear, planar and volume implants are studied. Results suggest that this anisotropy should be taken into account especially when one is considering dose distribution near the seeds. Angularly averaged distribution of single seed is found to cause serious deflections to isodoses of the implant in high dose region. To use true anisotropic dose rate distribution one must be able to evaluate angular orientation of seeds in the implant. A method is presented to compute this orientation also when seeds are poorly visible in radiographs. (orig.)
Abstract (German)
Die Dosisleistungsverteilung handelsueblicher I-125-Seeds ist aufgrund von Eigenabsorption stark anisotrop. Die Auswirkungen dieser Anisotropie auf die Dosisverteilung linearer, ebener und raeumlicher Implantate werden untersucht. Die Ergebnisse zeigen, dass diese Anisotropie insbesondere im Hinblick auf die Dosisverteilung in unmittelbarer Naehe der Seeds beruecksichtigt werden sollte. Es zeigt sich, dass die Annahme von Mittelwerten fuer die Winkelausrichtung der einzelnen Seeds zu starken Abweichungen der Isodosen des Implantats in dem mit hoher Dosis bestrahlten Gebiet fuehrt. Die Anwendung einer wirklich anisotropen Dosisleistungsverteilung setzt die Moeglichkeit voraus, die Ausrichtung der Seeds im Implantat festzustellen. Hierfuer wird eine Berechnungsmethode vorgestellt, die auch dann Anwendung finden kann, wenn die Seeds im Roentgenbild nur schwach zu erkennen sind. (orig.)Additional details
Publishing Information
- Journal Title
- Strahlentherapie und Onkologie
- Journal Volume
- 165
- Journal Issue
- 5
- Series
- Strahlenther. Onkol.
- Journal Page Range
- 396-400
- ISSN
- 0179-7158
- CODEN
- STONE
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 20060520
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ANISOTROPY; DOSE RATES; IODINE 125; ISODOSE CURVES; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIATION SOURCE IMPLANTS; RADIOTHERAPY; SELF-ABSORPTION
- Descriptors DEC
- ABSORPTION; BETA DECAY RADIOISOTOPES; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; IMPLANTS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; IODINE ISOTOPES; ISOTOPES; MEDICINE; NUCLEI; ODD-EVEN NUCLEI; RADIATION SOURCES; RADIOISOTOPES; THERAPY