Published April 1991 | Version v1
Report Open

Calibration of a PIXE system for analysis of environmental aerosols

  • 1. ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

Description

The results of the calibration of the PIXE system for analysis of environmental aerosols in the CNM is presented. The system was calibrated in order to lead to the project study of environmental aerosols using PIXE from the qualitative dimension to the quantitative one. Part of this work it was carried out in collaboration with the expert of the IAEA Dr. D. Crumpton. The calibration of the system it was achieved bombarding thin targets of well-known composition with protons of 2.5 MeV. Using this technique it was determined the absolute elementary concentration in samples of atmospheric aerosols. As an example the calculation of the lead concentration in one of them whose result was of 662 ± 49 ng m-3 is presented. This value agrees with the measurements made in the Crocker Nuclear Lab of the UCD. Its are also discussed the reproducibility and precision of the used technique. (Author)

Availability note (English)

Available from INIS in electronic form

Files

38058875.pdf

Files (509.6 kB)

Name Size Download all
md5:4aad0aab85dfac7613135d0a5ebe2795
509.6 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Calibracion de un sistema PIXE para analisis de aerosoles ambientales

Publishing Information

Imprint Pagination
12 p.
Report number
ACEL--9104

Optional Information

Secondary number(s)
INIS-MX-RI--0264