Implementation of DMLC quality control using EPID (Portal Dosimetry)

  • 1. Universidade de Sao Paulo (INRAD/HC/FMUSP), Sao Paulo, SP (Brazil). Instituto de Radiologia. Setor de Radioterapia
  • 2. Universidade de Sao Paulo (USP), Sao Paulo, SP (Brazil). Faculdade de Medicina

Description

A Quality Assurance (QA) to ensure the expected performance of a Multileaf Collimator System (MLC) is essential to deliver dose in a safety and appropriate way. The time required for equipment control and dosimetry may be reduced when the Electronic Portal Image Device (EPID) is used. The aim of this work was to check the resolution limits of the detection system for IMRT mode, and to propose a set of tests that can provide positioning analysis of a multileaf system. A Varian iX Clinac equipped with an 80 leaf Millenium MLC, and an amorphous silicon based EPID (aS1000) was used. The EPID proved itself effective for detecting errors up to 0.5 mm. The proposed tests provided relevant results of leaf position, and revealed that the MLC system is within acceptable limits found in literature. (author)

Availability note (English)

Available from http://www.rbfm.org.br/rbfm/article/view/371/v11n1p21

Additional details