Performance of Tantalum STJ X-ray Detectors at Elevated Count Rates

  • 1. STAR Cryoelectronics (United States)
  • 2. Lawrence Livermore National Laboratory (United States)

Description

We have developed an X-ray detector system based on Ta superconducting tunnel junctions (STJs) with an energy resolution below 10 eV under 1 keV. A principal advantage of STJs over other low-temperature detectors is their high count rate capability. In this work, we have optimized the operating conditions and digital pulse processing parameters for input count rates up to 25 kcps/pixel. Resolution below 10 eV is maintained up to 2.5 kcps/pixel, and at input rates of 25 kcps per pixel the resolution is 26 eV FWHM at 525 eV.

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Low Temperature Physics
Journal Volume
194
Journal Issue
1-2
Journal Page Range
p. 136-141
ISSN
0022-2291
CODEN
JLTPAC

Optional Information

Copyright
Copyright (c) 2019 Springer Science+Business Media, LLC, part of Springer Nature