200-eV resolution Si(Li) detector X-ray system based on pulsed opto-feed back technique
Creators
- 1. Bhabha Atomic Research Centre, Bombay (India). Nuclear Physics Div.
Description
The energy resolution of Si(Li) detector X-ray spectrometer has been improved to 200 eV by adopting an improved FET mounting configuration and by using a selected FET. The count-rate performance of the system has been evaluated for input count-rates upto 90,000/sec at 8.0 and 16.0 μsec peaking time of amplifier shaping network. The deterioration of resolution upto 30,000/sec input count-rates is insignificant. The peak stability is excellent and the peak shift observed upto 20,000/sec is found to be less than 5.0 eV. (M.G.B.)
Additional details
Publishing Information
- Publisher
- Department of Atomic Energy.
- Imprint Place
- Bombay (India)
- Imprint Title
- Proceedings of the nuclear physics and solid state physics symposium [held at] Madras, December 26-30, 1979
- Imprint Pagination
- 336 p.
- Journal Page Range
- p. 327-329.
Conference
- Title
- Nuclear physics and solid state physics symposium.
- Dates
- 26-30 Dec 1979.
- Place
- Madras (India).
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 14784119
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COUNTING RATES; ELECTRONIC CIRCUITS; ENERGY RESOLUTION; FIELD EFFECT TRANSISTORS; IRON 55; LI-DRIFTED SI DETECTORS; PREAMPLIFIERS; PULSE SHAPERS; X-RAY SPECTROMETERS
- Descriptors DEC
- AMPLIFIERS; BETA DECAY RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; ELECTRONIC EQUIPMENT; EQUIPMENT; EVEN-ODD NUCLEI; INTERMEDIATE MASS NUCLEI; IRON ISOTOPES; ISOTOPES; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; NUCLEI; PULSE CIRCUITS; RADIATION DETECTORS; RADIOISOTOPES; RESOLUTION; SEMICONDUCTOR DETECTORS; SEMICONDUCTOR DEVICES; SI SEMICONDUCTOR DETECTORS; SPECTROMETERS; TRANSISTORS; YEARS LIVING RADIOISOTOPES