Published January 1, 2017 | Version v1
Journal article

Innovative detectors for quality assurance dosimetry in SBRT of stationary and movable targets

  • 1. Centre for Medical Radiation Physics, University of Wollongong (Australia)
  • 2. North Shore Cancer Care Centre (Australia)
  • 3. Department of Radiation Oncology, Prince of Wales Hospital (Australia)
  • 4. SPA-BIT, Kiev (Ukraine)

Description

The high spatial and temporal resolution 2D monolithic silicon detector arrays M512 and DUO for quality assurance (QA) in real time motion adaptive radiotherapy (ART) have been developed. The DUO array possesses a spatial resolution of 0.2 mm and has demonstrated agreement within 5% with EBT3 film measurements of 6MV linac beam profiles for field sizes 1 × 1 cm2 and SRS cone diameter 0.5 cm. Dynamic characterisation of the M512 for QA in real time ART evaluated the performance of M512 for small fields while the detector experiences periodic motion. It was demonstrated with M512 that MLC tracking with Calypso electromagnetic array compensates for the periodic motion and improves the agreement between static and dynamic beam profiles for field size 1 × 1 cm2 from within 75% in the penumbra to within 11% agreement. The dynamic profile is returned to a similar distribution as the static case. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/777/1/012014

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
777
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
Workshop on micro-mini and nano-dosimetry and innovative technologies in radiation therapy
Acronym
MMND ITRO 2016
Dates
26-28 Jan 2016
Place
Tasmania (Australia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
49008074
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BEAM DYNAMICS; BEAM PROFILES; DOSIMETRY; FILMS; LINEAR ACCELERATORS; PERFORMANCE; PERIODICITY; QUALITY ASSURANCE; RADIOTHERAPY; SI SEMICONDUCTOR DETECTORS; SILICON; SPATIAL RESOLUTION
Descriptors DEC
ACCELERATORS; DYNAMICS; ELEMENTS; MEASURING INSTRUMENTS; MECHANICS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; RESOLUTION; SEMICONDUCTOR DETECTORS; SEMIMETALS; THERAPY; VARIATIONS