Published March 1994 | Version v1
Journal article

Automatic treatment method of background continuum of γ-ray spectra from whole body counting with NaI(Tl)

  • 1. Beijing Inst. of Radiation Medicine (China)

Description

The authors describe a computer method of automatic generation of background continuum from NaI(Tl) whole-body counting Gamma-ray spectra, according to principle of digital interpolation and etch peak. This simulative background was used to take the place of the background measured from 'background person'. The computer program of the method has been written in BASIC and FORTRAN-77, and run and tested on the TRS-80 and COPAM-PC-286S computer respectively. A complete whole-body spectra analysis program has been made up from combination of this program ATBK and the ND-ASAP software through a transform code READ written in the laboratory

Additional details

Publishing Information

Journal Title
Radiation Protection (Taiyuan)
Journal Volume
14
Journal Issue
2
Journal Page Range
p. 148-153.
ISSN
1000-8187
CODEN
FUFAEM