Published August 2007 | Version v1
Journal article

Poly (3-(3-Pyridyl) Acrylic Acid) Modified Glassy Carbon Electrode for Simultaneous Determination of Dopamine, Ascorbic Acid and Uric Acid

  • 1. College of Chemistry and Materials Science, Anhui Key Laboratory of Functional Molecular Solids, Anhui Normal University, Wuhu (Switzerland)

Description

Trans-3-(3-pyridyl) acrylic acid (PAA) was deposited on glassy carbon electrode (GCE) by electro polymerization in pH 7.0 phosphate buffer solution (PBS). The poly (3-(3-pyridyl) acrylic acid) (PPAA) film modified glassy carbon electrode shows an excellent electrochemical response for dopamine (DA), ascorbic acid (AA) and uric acid (UA). The cyclic voltammetry oxidation peaks for DA and AA, DA and UA, AA and UA are separated by 150 mV, 130 mV and 280 mV, respectively. This permits the simultaneous determination of AA, DA and UA. The interference of AA with the determination of DA could be eliminated because of the electrostatic interaction between DA cations and the negatively charged PPAA film at pH 7.0. The anodic peak currents of DA, AA and UA increase linearly with concentration in the range of 1-40 μmol L-1, 10-400 μmol L-1 and 1.6-80 μmol L-1, respectively, with a correlation coefficient (r) always higher than 0.998. The detection limit is 0.06 μmol L-1, 0.8 μmol L-1 and 1.1 μmol L-1 for DA, AA and UA, respectively

Additional details

Publishing Information

Journal Title
Annali di Chimica (Rome)
Journal Volume
8
Journal Page Range
p. 665-674
ISSN
0003-4592