Interpretation of the readings of combined albedo-track personnel neutron dosemeters in the terms of dose equivalent

  • 1. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Serpukhov. Inst. Fiziki Vysokikh Ehnergij

Description

Various perspective combined albedo-track personnel dosemeters with threshold (238U, 232Th and 237Np) and albedo (235U and 6LiF) neutron detectors have been discussed. For interpretation of the readings of combined dosemeters in terms of dose equivalent the matrix method based on the calculational analysis of the detector responses in various neutron spectra have been used. Calculated data have been compared with experimental ones. Perspectives of practical use of various combined dosemeters in personnel monitoring have been discussed. (author)