Helical tomotherapy shielding calculation for an existing LINAC treatment room: sample calculation and cautions
Creators
- 1. Department of Radiation Oncology, University of California-Davis, 4501 X Street, Sacramento, CA 95817 (United States)
Description
This paper reports a step-by-step shielding calculation recipe for a helical tomotherapy unit (TomoTherapy Inc., Madison, WI, USA), recently installed in an existing Varian 600C treatment room. Both primary and secondary radiations (leakage and scatter) are explicitly considered. A typical patient load is assumed. Use factor is calculated based on an analytical formula derived from the tomotherapy rotational beam delivery geometry. Leakage and scatter are included in the calculation based on corresponding measurement data as documented by TomoTherapy Inc. Our calculation result shows that, except for a small area by the therapists' console, most of the existing Varian 600C shielding is sufficient for the new tomotherapy unit. This work cautions other institutions facing the similar situation, where an HT unit is considered for an existing LINAC treatment room, more secondary shielding might be considered at some locations, due to the significantly increased secondary shielding requirement by HT. (note)
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/51/N389/pmb6_21_n02.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/51/N389/pmb6_21_n02.pdf; http://www.iop.org/;
- DOI
- 10.1088/0031-9155/51/21/N02;
- PII
- S0031-9155(06)23500-0;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 51
- Journal Issue
- 21
- Journal Page Range
- p. N389-N392
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38008660
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; LINEAR ACCELERATORS; PATIENTS; RADIOTHERAPY; SHIELDING
- Descriptors DEC
- ACCELERATORS; DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY; TOMOGRAPHY