Published September 11, 1998 | Version v1
Journal article

The anisotropy coefficient in Cerenkov counting

Description

We present an improved Cerenkov standardization method, which takes into account the directional nature of the Cerenkov light. A new expression for the Cerenkov counting efficiency is derived by assuming that the generation of photoelectrons at the photocathode is described by the Poisson law. In this procedure, the average energy that Cerenkov light needs to create one photoelectron at the photocathode and the anisotropy coefficient are confirmed as essential parameters for the characterization of the detection system. We study the influence of different measurement conditions (i.e., vial type, volume, acid and salt concentrations or spectrometer model) on these two parameters

Additional details

Identifiers

PII
S0969804397100185;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
49
Journal Issue
9-11
Journal Page Range
p. 1049-1053
ISSN
0969-8043
CODEN
ARISEF

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34013533
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
ANISOTROPY; CHERENKOV COUNTING; CONCENTRATION RATIO; EFFICIENCY; PHOTOELECTRIC EMISSION; STANDARDIZATION; VOLUME
Descriptors DEC
COUNTING TECHNIQUES; ELECTRON EMISSION; EMISSION; PHOTOELECTRIC EFFECT

Optional Information

Copyright