Published May 2015 | Version v1
Journal article

Usability evaluation through gonad shielding production of pediatric patients by gender and age rating

  • 1. Dept. of Radiology, Kyung Hee University Hospital at Gangdong, Seoul (Korea, Republic of)
  • 2. Dept. of Nuclear Engineering, Chosun University, Gwangju (Korea, Republic of)
  • 3. Dept. of Radiological Technology, Gwangju Health University, Gwangju (Korea, Republic of)

Description

The gonad shielding is used to minimize the impact of the exposure to gonads when Abdomen, Plevis and Hip X-ray inspections are conducted on radiation impressionable pediatric patients. By the way, the gonad is palpable difficult and impossible to check visually because it's a sensitive area, so tests are conducted with the approximate location of shielding, thereby appearing problems of not shielding gonads accurately. Accordingly, this study produced shields by age and gender of pediatric patients and studied the method of positioning shields with ASIS as a reference point without palpable sensitive areas, and tried to evaluate its usability. The study surveyed 30 pediatric patients by gender and age, who came and got inspected in Department of Radiology, our hospital from February 2012 to January 2014 and obtained the value of tolerance by measuring the average size of the pelvis using the distance measurement function of Infinitt Piview with the images stored in the PACS and producing shields by age and gender of pediatric patients and specifying the areas at random for the comparative analysis of pre- and post-using. It calculated the technology statistics (mean±SD) with the value of tolerance measured the length using SPSS 12.0 statistical program. As for boys, differences in the tolerance range of pre- and post-using shields were 2.69 mm in case of 1 year old, 2.58 mm in 2 years, 2.37 mm in 3 years, 2.815 mm in 4-5 years, 2.043 mm in 7-10 years, and as for girls,1.92 mm in 1-2 years, 1.75 mm in 3-4 years, 2.52 mm in 5-6 years and 1.93 mm in 7-10. After analyzing the pre- and post-using shields for all of boys and girls, there were statistically significant differences (P<0.050). It is considered that we can minimize the exposure to gonads and get a better video for diagnosis in testing high biological impressionable pediatric, if we use shields correctly with ASIS as a reference point considering its shape and size by age and gender in Abdomen, Plevis and Hip X-ray inspections

Additional details

Publishing Information

Journal Title
Journal of Radiation Industry
Journal Volume
9
Journal Issue
2
Series
8 refs, 5 figs, 4 tabs
Journal Page Range
p. 69-75
ISSN
1976-2402

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
47084812
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ABDOMEN; ACCURACY; DIAGNOSIS; LEGS; PATIENTS; PEDIATRICS; PELVIS; RADIATION PROTECTION; SHIELDING
Descriptors DEC
BODY; LIMBS; MEDICINE