Published 1978
| Version v1
Report
Restricted
3He detector design for low-level transuranic waste assay
Description
From this study it appears that a logical configuration of 3He detectors imbedded in CH2 for nondestructive assay of low-level transuranic waste is: rho = 1 atmosphere, d = 2.54 cm, x = 5.08 cm, t = 15.24 cm, h = 61 cm, and N = 5. If a greater detector response is desired, the best way to achieve it is to first double the detector height, h, and then to increase the number of tubes, N, and/or increase the 3He density, rho
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.
Files
Additional details
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- LA-UR--78-326
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 11506049
- Subject category
- S12: MANAGEMENT OF RADIOACTIVE WASTES, AND NON-RADIOACTIVE WASTES FROM NUCLEAR FACILITIES; S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- HE-3 COUNTERS; LOW-LEVEL RADIOACTIVE WASTES; NONDESTRUCTIVE ANALYSIS; POLYETHYLENES; SPECIFICATIONS