Volume recombination of the induced ionization in an X-ray ionization chamber is an important factor affecting the collection efficiency of the charge when such chambers are operated at atmospheric pressure. The volume recombination process is also dependent on the X-ray exposure rate. The theory for recombination in a cylindrical ionization chamber is shown to be in agreement with experimental measurements. For the MDH 10X5-6 cylindrical ionization chamber, the recombination loss is unimportant at exposure rates consistent with its intended usage. (U.K.)
Additional details
Publishing Information
- Journal Title
- Radiat. Prot. Dosim.
- Journal Volume
- 3
- Journal Issue
- 3
- Series
- Radiat. Prot. Dosim.
- Journal Page Range
- 175-179
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 14780245
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CORRECTIONS; IONIZATION CHAMBERS; RECOMBINATION; X-RAY DOSIMETRY
- Descriptors DEC
- DOSIMETRY; MEASURING INSTRUMENTS; RADIATION DETECTORS