Published 1982 | Version v1
Journal article

Ion recombination characteristics of the MDH 10X5-6 ionisation chamber under continuous exposure

Creators

  • 1. Bureau of Radiological Health, Rockville, MD (USA). Div. of Electronic Products

Description

Volume recombination of the induced ionization in an X-ray ionization chamber is an important factor affecting the collection efficiency of the charge when such chambers are operated at atmospheric pressure. The volume recombination process is also dependent on the X-ray exposure rate. The theory for recombination in a cylindrical ionization chamber is shown to be in agreement with experimental measurements. For the MDH 10X5-6 cylindrical ionization chamber, the recombination loss is unimportant at exposure rates consistent with its intended usage. (U.K.)

Additional details

Publishing Information

Journal Title
Radiat. Prot. Dosim.
Journal Volume
3
Journal Issue
3
Series
Radiat. Prot. Dosim.
Journal Page Range
175-179
ISSN
0144-8420

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
14780245
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CORRECTIONS; IONIZATION CHAMBERS; RECOMBINATION; X-RAY DOSIMETRY
Descriptors DEC
DOSIMETRY; MEASURING INSTRUMENTS; RADIATION DETECTORS