Calculation of the detection efficiency in liquid scintillators. III. Single electron-capture radionuclides

Description

In this paper, counting efficiency vs figure of merit for 19 electron capture radionuclides has been computed. Toluene based s c i n t i l l a t o r solution of 15 cm3 has been assumed in the computation. The relationships to obtain the probability per decay are given and the computed efficiency tabulated.(Author)